2.3 Flash measurement (only DIGIFLASH 2)

2.3.1 Setting the sync speed (measuring time)

Press the Function button for selecting the

function „flash measure-

ment“. A small stroke in front of the sync time will appear in the
display

to signal „flash metering“.

Keep the Function button pressed down until the ok signal sounds.
The display will now show the sync speed blinking.

Use the measuring key to input the desired sync speed, in the range
of 1 sec to1/500 sec. and confirm by pressing the function button.


2.3.2 Measuring

Push the measuring key. The DIGIFLASH 2 is ready to measure

for

30 sec, as long as „F“ in the display is on.

Trigger the flash.

Transfer the value measured from the display to the window in

the printed scales above and set it precisely under the indicator
line.

Read the f/stop in the upper part of the scale opposite the sync

speed you had preselected (for 1/90 read between 1/60 and 1/125).