Electromagnetic immunity – Welch Allyn CP150 - User Manual Benutzerhandbuch

Seite 50

Advertising
background image

Emissions test

Compliance

Electromagnetic environment - guidance

IEC 61000-3-2

Voltage fluctuations/

flicker emissions
IEC 61000-3-3

Complies

Electromagnetic immunity

The CP 150 electrocardiograph is intended for use in the electromagnetic environment

specified below. The customer or user of the CP 150 electrocardiograph should assure

that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic

environment -

guidance

Electrostatic discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood,

concrete, or ceramic tile. If

floors are covered with

synthetic material, the

relative humidity should be

at least 30%.

Electrical fast transient/burst
IEC 61000-4-4

±2 kV for power supply lines
±1 kV for input/output lines

±2 kV for power supply lines
Not applicable

Mains power quality should

be that of a typical

commercial or hospital

environment.

Surge
IEC 61000-4-5

±1 kV differential mode
±2 kV common mode

±1 kV differential mode
±2 kV common mode

Mains power quality should

be that of a typical

commercial or hospital

environment.

Voltage dips, short

interruptions, and voltage

variations on power supply

input lines.
IEC 61000-4-11

>95% dip for 0.5 cycle
60% dip for 5 cycles
30% dip for 25 cycles
>95% dip for 5 seconds

>95% dip for 0.5 cycle
60% dip for 5 cycles
30% dip for 25 cycles
>95% dip for 5 seconds

Mains power quality should

be that of a typical

commercial or hospital

environment. If the user of

the CP 150

electrocardiograph requires

continued operation during

power mains interruptions, it

is recommended that the CP

150 electrocardiograph be

powered from an

uninterruptible power supply

or battery.

Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical

commercial or hospital

environment.

46 EMC guidance and manufacturer’s declarations

CP 150 Ruhe-Elektrokardiograph mit 12 Ableitungen