Recommended separation distances between portable

and mobile RF communications equipment and the CP 150

electrocardiograph

The CP 150 electrocardiograph is intended for use in an electromagnetic environment in

which radiated RF disturbances are controlled. The customer or user of the CP 150

electrocardiograph can help prevent electromagnetic interference by maintaining a

minimum distance between portable and mobile RF communications equipment

(transmitters) and the CP 150 electrocardiograph as recommended below, according to

the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter (m)

Rated max. output power of

transmitter

(W)

150 kHz to 80 MHz

d = (1.2)

80 MHz to 800 MHz

d = (1.2)

800 MHz to 2.5 GHz

d = (2.3)

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended

separation distance d in meters (m) can be estimated using the equation applicable to

the frequency of the transmitter, where P is the maximum output power rating of the

transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range

applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects, and people.

48 EMC guidance and manufacturer’s declarations