ALTANA Temp-Gard Basic Benutzerhandbuch
Seite 35
temp-gard basic Hardware
14
Starting a Measurement
• Connect the thermocouple probes to the temp-gard basic instrument.
• Insert an SRAM card with a standard template into the slot of the temp-gard basic
instrument.
• Press
the
start/stop key to switch on the instrument. The template setup is displayed.
• The preset number of probes, the duration and the sampling interval of the measure-
ment can be adjusted by pressing the
mode key of the temp-gard basic instrument.
• Start the measurement by pressing the start/stop key.
• Embed
the
temp-gard basic in the thermal barrier.
Attention
Make sure that the thermal barrier have cooled down completely after the previous
measurement.
• Screw down the heat sinks (optional) in both halfs of the thermal barrier.
• Place
the
temp-gard basic in the lower part of the thermal barrier.
• Make sure that all gaskets are clean and without defects.
• Place the thermocouple cables side by side in the lateral sparings of the thermal
barrier and lock the box with the upper part.
Recording the Oven Temperature
• Attach the thermocouple probes to the object.
Attention
Secure all cables and make sure that they cannot get tangled up when travelling through
the oven.
• Hang the thermal barrier on the belt and make sure there is enough free space for the
thermocouple probes, the cables and the thermal barrier.
Finishing the Measurement
Attention
All parts can be hot after the measurement.
Please observe the safety information on page 3.
!
!
!