Description of the analog I/O interface

The

analog I/O interfaces are accessories for VACUUBRAND

-

Controller CVC 3000 from software version V 1.00 (configuration from software version V 1.40, see

section “Configuration of the interfaces”)

-

Controller VNC 2 from software version V 1.21

-

Measurement device DCP 3000 from software version V 1.00

The analog I/O interface is designated to be connected to the VACUU•BUS

TM

female plug (with exten-

sion cable VACUU•BUS

TM

).

Connect the analog I/O interface only if the device is shut off.

When switching on the device the analyzable analog I/O interfaces are detected automatically, depend-

ing on controller / measurement device. Not analyzable interfaces are ignored.

Each analog I/O interface needs an individual BUS address (see section “Configuration of the inter-

faces”). Do not use interfaces with the same address within one BUS system. Configuration of the inter-

faces is only possible with the controller CVC 3000.

At the

CVC 3000 the interfaces can be used for

- output of vacuum as analog voltage 0-10 V (factory set)

- input of set vacuum and output of actual vacuum as analog voltage 0-10 V

- output of motor speed as analog voltage 0-10 V

- input and output of motor speed as analog voltage 0-10 V

At the

VNC 2 the interfaces can be used for

- output of vacuum as analog voltage 0-10 V (factory set)

- input and output of vacuum as analog voltage 0-10 V (only after configuration with a vacuum control-

ler CVC 3000)

At the

DCP 3000 the interfaces can only be used in the delivery status as output of vacuum as analog

voltage 0-10 V.

Error messages when removing (pull off at the controller) an analog I/O interface

CVC 3000: Interfaces configured for input and output of vacuum or motor speed lead to an error mes-

sage.

VNC 2: Interfaces configured for input and output of vacuum lead to an error message.

The error message due to a removed analog I/O interface has to be reset by loading the factory setting

at the controller.

Attention: If loading the factory setting at the controller CVC 3000 all programs are deleted!