Function of the digital I/O interface

Factory-set the digital I/O interface is configurated as

fault indicator for internal faults (e. g. valve,

VARIO pump, sensor failure, ....) as well as for external faults (apparatus / application of the user).

The digital I/O interface has a galvanically isolated switching input and switching output.

The LED indicates via change of the colour the switching status of the input and the output and therefore

the kind of error:

The LED indicates the switching status of the input and the output:

- green: no fault, signal at the input correct (5-75 VDC or 5-50 VAC), output switched (switching out-

put closed).

- yellow: internal fault (valve, VARIO pump, sensor failure, ....), signal at the input correct (5-75 VDC

or 5-50 VAC), output not switched (switching contact open).

- red: external fault: signal at the input incorrect (< 5 VDC or 5 VAC), output not switched (switching

contact open).

Error messages have to be quit (deleted) after elimination of the fault by pressing key START/STOP at

the CVC 3000.

When using the digital interface with the controller CVC 3000 or VNC 2 or the measurement de-

vice DCP 3000 a signal (5-75 VDC oder 5-50 VAC) at the input is necessary. Otherwise vacuum

control is stopped and / or the device indicates a fault (flashing warning triangle).

Only if using the CVC 3000 or the DCP 3000 the digital I/O interface can be configurated for function

“Remote (Start - Stop)” or “End” or „Valve“ (Isol.v. 1 2 3 4, Waterv. 1 2 3 4, Air. V. 1 2 3 4, VMS 1 2 3).

Notes concerning the configuration see section “Configuration of the interfaces).

-

Remote module (Start - Stop): Pressure control can be started with a signal (5-75 VDC or 5-50

VAC) at the input of the interface. The output contact indicates the status of the control (switching

contact closed: control is running). Control stops if there is no voltage signal at the input (output

switching contact closed).

The LED indicates the operation status: red indicates Stop, green indicates Start.

-

End module: In this configuration no signal at the input is necessary. At the output e. g. a valve ( e.

g. venting valve) or a relais is connected which opens if process control is stopped.

-

Valve adapter module: The digital I/O interface can be used as interface for exsisting (provided by

the customer) isolation, coolant or venting valves or a VACUUBRAND VMS module, if these com-

ponents are not equipped with a VACUU•BUS

TM

connection. The component has to be connected to

the output of the module (an additional voltage supply is necessary).