Das Dokument darf nur vollständig und unverändert verwendet und weitergegeben werden. Es liegt in der Verantwortung des

Anwenders, die Gültigkeit dieses Dokumentes bezüglich seines Produktes sicher zu stellen. BA-Nr.: 999236 / 05/09/2011

Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999236 / 05/09/2011

Technical Data

We reserve the right for technical modification without prior notice!

Replacing the device fuse

Switch off the device and disconnect it from mains before replacing the fuse. After

disconnecting from mains wait two minutes to allow the capacitors to discharge.

Identify and eliminate the cause of failure before switching on the device again.

Keep the snap-fit squeezed and pull the fuse holder out.

The fuse holder contains two fuses of the same type. Replace the defective fuse

by a fuse of the same type (see ”Technical data”).

Insert the fuse holder into the device until it snaps.

fuse holder

Type

VMS-B-Module

Mains connection

100-230 V, 50-60 Hz

Max. power / current: 24V /VACUU•BUS

TM

1.1A

Nominal current

0.8 A

Degree of protection IP

20

Max. permitted ambient temperature during operation

+10°C to +40°C

Max. permitted ambient temperature during storage

-10°C to +60°C

Permitted relative atmospheric moisture during operation

(no condensation)

30% to 85%

Max. permitted power draw at connection vacuum pump for

ohmic (inductive) load

8(4) A bei 250 V~ / 8(6) A bei 125 V~

Fuse

2 x 250 V

Dimensions

179 mm x 108 mm x 61 mm

Mass

0.7 kg