SCPI-Befehlsreferenz

364

FTB-5240S/S-P/BP

Product-Specific Commands—Description

Product-Specific Commands—Description

:ABORt

Description

This command resets the trigger system and

places all trigger sequences in the IDLE state. Any

trace acquisition that is in progress is aborted as

quickly as possible. The command is not

completed until the trigger sequence is in the

IDLE state.

This command is an event and has no associated

*RST condition or query form.

Syntax

:ABORt

Parameter(s)

None

Example(s)

ABOR

Notes

A call to ABORt only returns once acquisition is

completely stopped and instrument is ready for

new commands. For this reason, execution of

this command may take a few seconds.

For a continuously initiated acquisition

(INIT:CONT ON), calling ABORt will

automatically set it to OFF.

See Also

:INITiate[:IMMediate]

:INITiate:CONTinuous

:STATus

:STATus:OPERation:BIT<8|9>:CONDition?