Technical data
LF413T
14
ba55324e03
10/2009
4
Technical data
General features
Measuring principle
Four-electrode measurement
Cell constant
0.475 cm
-1
±1.5 %
Temperature sensor
integrated NTC 30 (30 k
Ω
/ 25 °C)
Dimensions
(in mm)
Weight
approx. 135 g
Materials
Shaft
Epoxy
Connection head
POM
Conductivity electrodes
Graphite
Thermistor enclosure
Graphite
Connection cable
Length
3 or 1,5 m
Diameter
6 mm
Smallest allowed
bend radius
Permanent bend:
80 mm
Single time or short time bend: 50 mm
Plug type
Socket, 8 pins
Pressure
resistance
Sensor with closed plug
connection
IP 68 (2 x 10
5
Pa or 2 bar)
Cable plug
IP 67 (when plugged in)
The LF413T meets the requirements according to article 3(3) of the
97/23/EC directive ("Pressure equipment directive").