Xylem LF413T_LF413T-ID Benutzerhandbuch

Seite 15

Advertising
background image

LF413T

Technical data

15

ba55324e03

10/2009

Pin assignment

Measurement

conditions

Conductivity measuring
range

1

μ

S/cm ... 2 S/cm

Temperature range

-5 ... 80 °C (100 °C)

Max. allowed overpressure

2 x 10

5

Pa (2 bar)

Minumum depth of
immersion

36 mm

Maximum depth of
immersion

Entire sensor+cable (up to 80 °C)
Sensor shaft only (=120 mm / up to 100 °C)

Operating position

any

Storage

conditions

Recommended storing
method

in air

Storage temperature

0 ... 50 °C

Characteristic data

on delivery

Temperature responding
behavior

t

99

(99 % of the final value after) < 20 s

Precision of the temperature
sensor

± 0.2 K

4

2

5

3

8

6

1

7

NTC
30 k

W

Voltage electrode 2

Current electrode 1

Voltage electrode 1

NTC

NTC

Current electrode 2

Shield

not used

Plug from the front:

4

5

7

3

8

2

6

1

Advertising