Calibration

Calibrate according to the data given in the

operating manual of the measuring instrument and

the analysis specification.

Sample preparation

Add 2 % ISA/FK solution. This sample conditioning

solution creates optimum conditions for measuring. It

provides a constant ionic strength and similar diffusion

potentials at the reference electrode in standard and test

sample.

If you would like to have more detailed information

concerning sample preparation and measuring

procedures, SI Analytics provides a large number of

application reports for various applications.

Response times

The response time depends on the concentration range.

It is:

several seconds with high concentrations,

several minutes near the detection limit.

The measured value is stable if the value does not change
by more than 0.1 mV within 30 seconds.