Storage – Xylem ISE Electrodes I 60 Benutzerhandbuch

Seite 15

Advertising
background image

15

ACCESSORIES

). To do so, moisten a polishing

trip with deionized water, and with circular
movements polish the membrane surface for
approx. 30 seconds.

– Subsequently place the electrode into diluted

standard solution for approx. 5 minutes.

Storage

Between two measurements, place the

combination

electrode into diluted standard

solution.

Overnight up to one week: Push the closing ring

upwards to close the refilling opening. Then rinse

the combination electrode with deionized water,

blot it dry using a clean paper towel and store it

in a dry place.

For more than a week: Remove the electrolyte

solution from the reference electrode and rinse

the combination electrode with deionized water,

blot it dry using a clean paper towel and put on

the protection cap. Store the combination

electrode in a dry place.