Sample preparation, Before measuring – Xylem ISE Electrodes Ag-S 60 Benutzerhandbuch

Seite 15

Advertising
background image

15

Minimum depth of immersion:

The diaphragm must be covered

Maximum depth of immersion:

Approx. 1 cm below the fluid level of the bridge electrolyte

Before Measuring

x Before use, condition the combination electrode or

electrodes respectively for approx. 5 min. in standard
solution.

x Remove any air bubbles in the bridge electrolyte by

slightly knocking against the shaft.

x Calibrate according to the operating manual of the meter

and the analysis specification.

Sample preparation

Ag

+

measurement

Add 2 % ISA/FK solution. This sample conditioning solution

creates optimum conditions for measuring. It provides a constant
ionic strength and similar diffusion potentials at the reference

electrode in standard and test sample.

S

2-

measurement

Add 50 % SAOB II solution.

Preparation of the SAOB II solution:

x

200 ml/L 10 molar NaOH

x

35 g/L ascorbic acid

x

67 g/L EDTA

Note

A freshly prepared SAOB II solution (Sulfide Anti-Oxidant

Buffer) is colorless or slightly yellowish brown. If the solution
has already strongly oxidized its color is dark brown. In this
case it can no longer be used.

If you would like to have more detailed information
concerning sample preparation and measuring procedures,
SI Analytics provides a large number of application reports
for various applications.