Sample preparation

Add 2 % ISA/K solution. This sample conditioning
solution creates optimum conditions for measuring. It
provides a constant ionic strength and similar diffusion
potentials at the reference electrode in standard and test
sample.

If you would like to have more detailed information
concerning sample preparation and measuring
procedures, SI Analytics provides a large number of
application reports for various applications.

Response times

The response time depends on the concentration range.

It is:

several seconds with high concentrations,

several minutes near the detection limit.

The measured value is stable if the value does not

change by more than 0.1 mV within 30 seconds.

Interferences

Interfering ions: 10 % error with the following ratios of

concentration: (Ratio of concentration = Interfering ion /

Measuring ion)