Aging

Please note that every electrode undergoes a natural

aging process. The response time increases and the
slope decreases with the age of the electrode. The
following factors shorten the lifetime considerably:

• Incorrect storing
• Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of

interfering ions)

• High temperatures
• High changes in temperature

The warranty does not cover damage caused by
measuring conditions and mechanical damage.

Maintenance

•

For combination electrodes: replenish used reference

electrolyte solution.

• In case of increased response times or drifting

measured values, place the electrode into distilled
water for one hour and then store it in diluted
standard solution for several hours.

Storage

•

Between two measurements, place the
combination

electrode into diluted standard

solution.

• Overnight up to one week: Push the closing ring

upwards to close the refilling opening. Then rinse
the combination electrode with deionized water,
blot it dry using a clean paper towel and store it
in a dry place.