IMPAC pyrometers IN 300 · IN 300/5

21

9

Trouble shooting


Before sending the pyrometer for repair, try to find the error and to solve the problem with the help
of the following list.

Temperature indication too low
 Incorrect alignment of the pyrometer to the object

 New correct alignment to achieve the max. temperature signal

 Measuring object smaller than spot size

 check measuring distance, smallest spot size is at nominal measuring distance (see 5)

 Emissivity set too high

 Set lower correct emissivity corresponding to the material (see 6)

 Lens

contaminated

 Clean lens carefully (see 8.2)

Temperature indication too high
 Emissivity set too low

 Set lower correct emissivity corresponding to the material (see 6)

 The measurement is influenced by reflections of hot machine parts

 Use mechanical construction to avoid the influence of the interfering radiation

Measuring errors
 Indicated temperature is decreasing during the use of the pyrometer, contamination of the lens

 Clean lens. Recommendation: use of air purge

 Indicated temperature is decreasing during the use of the pyrometer, although the air purge

unit is used. Probably compressed air is not clean or air failed
 Clean the lens and use clean, dry and oil free compressed air

 HF-interferences

 Correct the connection of the cable shield (see 4)

 Instrument

overheated

 Use cooling jacket with air or water cooling