KROHNE UFM 3030 Service DE Benutzerhandbuch
Seite 16

16
UFM 3030
Main menu level
Main menu level
Submenu level
Submenu level
Function level
Function level
Data level
Data level
Fct. 4.00.00
PARAM ERR
Fct. 4.01.00
FLOW VELOC
Fct. 4.01.01
FULL SCALE
1.0000E+2
m3/hr
Fct. 1.00.00
OPERATION
Fct. 1.01.00
FLOW
Fct. 1.01.01
FULL SCALE
1.0000E+2
m3/hr
Fct. 1.00.00
STORE YES
Fct. 1.00.00
STORE NO
Fct. 1.00.00
RETURN
Fct. 4.00.00
STORE YES
Fct. 4.00.00
STORE NO
Fct. 4.00.00
RETURN
CODE 2
- -
100.0
m3/hr
Fct. 0.00.01
VIEW ERR
Error
messages
Data acceptance
level
(only appears
after change
of data)
Data acceptance
level
(only appears
after change
of data)
Measuring
mode
Measuring
mode
Error/
Totalizer reset
mode
Error/
Totalizer reset
mode
Parameter
setting
mode
Parameter
setting
mode
Parameter
error
mode
Parameter
error
mode
CODE 1
- - - - - - - - -
Parameter
check
9 x
START
START
Taste Messmodus
Menümodus
Datenebene
→
Zum Parametereinstellungsmodus
gehen, Funktion 1.00.00
OPERATION. Wenn Zugangs-CODE
1 aktiviert ist, muss zuerst CODE 1
eingegeben werden. Mit Funktion
3.07.02 kann Zugangs-CODE 1
aktiviert oder deaktiviert werden.
Zur nächsten, unteren
Menüebene gehen.
Zum nächsten Zeichen
oder zur nächsten Zeile
gehen (nur bei Anzeige
von mehr als einer
Zeile).
↵
Zum Modus Fehler-/Zähler-
Rücksetzung gehen (über "CODE 2").
Zur vorigen (höheren)
Menüebene
zurückkehren oder
Menümodus
verlassen.
Eingegebenen Wert
annehmen.
↑
Messwerte durchlaufen, siehe
Funktion 3.03.07 CYCL DISP.
Die Menüoptionen auf
der aktuellen
Menüebene
nacheinander
durchschalten.
Aktives Zeichen zu
neuen Werten
durchschalten.
Funktion
Text
Beschreibung und Einstellungen
0.00.00 ERROR/TOT
Hauptmenü 0.00.00 Error/Totalizer (Fehler/Zähler)
0.00.01
VIEW ERR
Ansicht der Liste der Fehlermeldungen
0.00.02
RST ERR
Fehlermeldungen zurücksetzen