Troubleshooting
- 25 -
Trouble
Possible Causes
Action to be taken
- Tip dripping (instrument leaks)
- Unsuitable tip
- Tip not seated tightly
- Only use high-quality tips
- Press tip on firmly
- Aspirated volume too low
- Pipette shaft cone loose
- Tighten clockwise see
“Handling”, page 8
- Delivered volume too low
- Seal contaminated
- see “Cleaning”, page 19
- Seal damaged
- see “Specifications/Spares”,
page 26
- Non-aqueous solutions pipetted
- see “Adjustment”, page 14
- Piston damaged
- see “Spares”, page 26
- Slow suction or none
- Shaft clogged
- Clean shaft
see “Cleaning”, page 19
- Volume too high
- Pipetting key pressed too far
(into the blow-out position)
before sample take up.
- Operate properly
see “Pipetting”, page 11