Host-kommunikation – Leica Biosystems LPC Software installation Benutzerhandbuch
Seite 182
182
Software-Installationshandbuch, Revision 2 - 03/2010
PICK RESULT
Value
Cassette pickup successful
0
Cassette pickup successful, magazine is low on cassettes
1
Pickup failed, out of cassettes
2
Pickup failed, latch opened
3
Pickup failed, marker is paused
4
Cannot pickup, exit bins are full
5
Pickup failed, general fault
6
Undefined. Marker has yet to pickup a cassette,
or a pickup operation is in progress.
-1
PRINT RESULT
Value
Printed OK
0
Aborted by E-Stop
1
Aborted by Operator
2
Aborted by printer timeout (took too long to print)
3
Aborted by printer reset command
4
Aborted by the printer
5
Aborted due to verify error
6
Undefined. Marker has yet to print a cassette,
or a print is in progress.
-1
Es ist Aufgabe des Host-Computers, in angemessenem Abstand Status-
informationen vom Producer CL abzufragen. Diese verändern sich stetig, da
das Gerät Zustände ändert und Kassetten verarbeitet, sich der Status all
dieser Felder also ändert. Tut das der Host-Computer zu unregelmäßig,
könnten Änderungen verpasst werden. Denn die Producer CL-Software
wird auf keinen Fall Statusänderungen in die Warteschlange stellen oder
aufbewahren.
4.