Software-Installationshandbuch, Revision 2 - 03/2010

PICK RESULT

Value

Cassette pickup successful

0

Cassette pickup successful, magazine is low on cassettes

1

Pickup failed, out of cassettes

2

Pickup failed, latch opened

3

Pickup failed, marker is paused

4

Cannot pickup, exit bins are full

5

Pickup failed, general fault

6

Undefined. Marker has yet to pickup a cassette,
or a pickup operation is in progress.

-1

PRINT RESULT

Value

Printed OK

0

Aborted by E-Stop

1

Aborted by Operator

2

Aborted by printer timeout (took too long to print)

3

Aborted by printer reset command

4

Aborted by the printer

5

Aborted due to verify error

6

Undefined. Marker has yet to print a cassette,
or a print is in progress.

-1

Es ist Aufgabe des Host-Computers, in angemessenem Abstand Status-
informationen vom Producer CL abzufragen. Diese verändern sich stetig, da
das Gerät Zustände ändert und Kassetten verarbeitet, sich der Status all
dieser Felder also ändert. Tut das der Host-Computer zu unregelmäßig,
könnten Änderungen verpasst werden. Denn die Producer CL-Software
wird auf keinen Fall Statusänderungen in die Warteschlange stellen oder
aufbewahren.