At the factory, the level sensor is adjusted either to the 4l polypropylene bottle or to the coated 2l glass
bottle (G version) at standard conditions.
It may be necessary to adjust the sensor to compensate for changes in bottle material, bottle size or
changed ambient conditions.
Adjustment of the BVC level sensor to work properly
with a new bottle, or to correct false alarms
Pertains only to the BVC professional / BVC professional G
NOTICE
Preparation
1. Do not switch off the BVC.
2. To shut off a false alarm prior to an adjustment, simply touch the key marked “bottle change.”
The red LED will continue to flash.
3. Empty the bottle.
4. Set the correct bottle size at the touch panel.
5. Wait approx. 1 minute before adjustment.
We also recommend emptying the bottle and adjusting the BVC professional if there is a false alarm
indicating “liquid level” even though the bottle is less than 3/4 full.
➨
If there is a false alarm in spite of an adjustment, repeat the adjustment.
➨
If there is a false alarm in spite of
repeated adjustment, please call +49 9342 808-5500 (after sales
service), or your local VACUBRAND sales office.
Do not attach adhesive foil or anything similar to level sen-
sor or to the side of the bottle directly next to the sensor.
adhesive foil
NOTICE
Use only the original VACUUBRAND BVC bottle or the original spare bottle (see
“Accessories - Spare parts”).