Adjustment routine 4 – VACUUBRAND BVC professional /G: Instruction d´ajustement le capteur de niveau de liquide Benutzerhandbuch
Seite 17

17/20
Action
Reaction
Meaning
Step 1
Touch “4l PP“ key.
The LED of the „4l PP“ bottle
symbol lights blue.
Ready for adjustment.
Step 2
Release “4l PP“ key.
Step 3
Touch and hold „2l Glass“ key. The LED of the „4l PP“ bottle
symbol turns off.
The LED of the „2l Glass“ bottle
symbol lights blue.
Adjustment is possible by con-
tinuing to hold the “2l Glass” key,
and proceeding to step 4.
Step 4
When the LED of the „2l Glass“
key lights blue, touch and hold
the “bottle change” key at the
same time.
Case 1:
A one-time buzzer sounds.
After the buzzer, the LED of the
“bottle change” key lights red.
Adjustment has been carried out
correctly.
Case 2 / false alarm:
The alert tone for the false alarm
is shut off.
A one-time buzzer sounds.
After the buzzer, the LED of the
“bottle change” key lights red.
Continue with step 5 and then
repeat adjustment after 1 minute.
Step 5
Release both „2l Glass“ key and
„bottle change“ key.
The LED of the “2l Glass” key
lights blue.
The red LED of the “bottle
change” key turns off.
Adjustment finished.
Adjustment routine 4
Attention: Do not interrupt the adjustment routine once you have begun, even if
there is an alarm. Complete the entire routine, and repeat if necessary.
Bottle
Software version
Serial number
Adjustment
routine
2l Glass
not specified
37040501- 37040502
4