Adjustment routine 2 – VACUUBRAND BVC professional /G: Instruction d´ajustement le capteur de niveau de liquide Benutzerhandbuch
Seite 15

15/20
Action
Reaction
Meaning
Step 1
Touch and hold “2l Glass” key.
Case 1:
The LED of the “2l Glass“ bottle
symbol flashes blue.
Adjustment is possible by
continuing to hold the “2l Glass”
key, and proceeding to step 2.
Case 2 / false alarm:
The alert tone of the false alarm
is shut off.
The LED of the “2l Glass“ bottle
symbol flashes blue.
Step 2
When the LED of the “2l Glass”
key flashes blue, touch and hold
the “bottle change” key at the
same time.
A one-time buzzer sounds.
After the buzzer, the LED of the
“bottle change“ key lights red.
Adjustment has been carried out
correctly.
Step 3
Release both “2l Glass” key and
“bottle change” key.
The LED of the “2l Glass“ key
lights blue.
The red LED of the “bottle
change” key turns off.
Adjustment finished.
Adjustment routine 2
Attention: Do not interrupt the adjustment routine once you have begun, even if
there is an alarm. Complete the entire routine, and repeat if necessary.
Bottle
Software version
Serial number
Adjustment
routine
2l Glass
V1.6 and higher
not relevant