Response times

The response time depends on the concentration range.

It is:

several seconds with high concentrations,

several minutes near the detection limit.

The measured value is stable if the value does not

change by more than 0.1 mV within 30 seconds.

Interferences

pH values < 5 or > 7

Aging

Please note that every electrode undergoes a natural
aging process. The response time increases and the
slope decreases with the age of the electrode. The
following factors shorten the lifetime considerably:

• Incorrect storing
• Special measuring conditions (e.g. organic solutions,

frequent measuring with high concentrations of
interfering ions)

• High temperatures
• High changes in temperature

The warranty does not cover damage caused by
measuring conditions and mechanical damage.