Maintenance, Storage – Xylem ISE Electrodes F 60 Benutzerhandbuch

Seite 19

Advertising
background image

19

Maintenance

• For combination electrodes: replenish used reference

electrolyte solution.

In case of increased response times of the electrode

perform the following activities:

- Polish the membrane surface using an SF/K
polishing strip. To do so, moisten a polishing trip
with deionized water, and with circular movements
polish the membrane surface for approx. 30
seconds.

- Subsequently place the electrode into diluted
standard solution for approx. 5 minutes.

Storage

Combination

electrode F 60

Between two measurements, place the
combination

electrode into diluted standard

solution.

Overnight up to one week: Push the closing ring
upwards to close the refilling opening. Then rinse
the combination electrode with deionized water,
blot it dry using a clean paper towel and store it
in a dry place.

For more than a week: Remove the electrolyte
solution from the reference electrode and rinse
the combination electrode with deionized water,
blot it dry using a clean paper towel and put on
the protection cap. Store the combination
electrode in a dry place.