6. Commissioning, operation

6.11 Measuring temperature with a resistance thermometer

For temperature measurement, select the function "RTD" on one of the displays.

The correct resistance thermometer must be used. There are four types of resistance

thermometer supported: "P100-385", "P100-392", "P100-JIS" and "CUSTOM".

The factory default type is "P100-385", so if the CPH6600 is being used

with a WIKA Pt100 you do not have to reset the resistance thermometer

type.

The WIKA Pt100 only needs to be connected to the CPH6600 and the measured

temperature will immediately be displayed.

The display will indicate "OL" if the measured temperature is outside

the nominal measurement range of the resistance thermometer (below

-40 °C or above 150 °C).

If a custom resistance thermometer is being used, the entry of R0 and the coefficients is

handled through the serial interface.

Figure 3: CPH6600 with a resistance thermometer