WIKA operating instructions, model CPH6600

14005103.03 11/2012 GB/D

GB

6. Commissioning, operation

8. Select the "NEXT" option to display the pressure at which the switch closed and the

hysteresis "SW DEADBAND".

9. Select the "NEW TEST" option to clear the data and perform another test.

10. Select the "DONE" option to end the test and return to the standard pressure

setting.

Example: [P1] ST will return to [P1].

The previous example uses a normally closed switch. The basic

procedure is effectively the same for a normally open switch, the display

will simply read "OPEN" instead of "CLOSE".

6.13 Calibrating transmitters
6.13.1 Using the mA measurement function

The mA function enables to read out the 4 ... 20 mA output from the instrument being

calibrated. This can be done in two ways.

1) Passively – the test item generates 4 ... 20 mA directly. The value is read on the

calibrator.

2) Actively – the calibrator supplies a voltage supply of DC 24 V to the test item to

power the device while reading the resulting 4 ... 20 mA signal.

Passive

Active