Metrohm 798 MPT Titrino Benutzerhandbuch

Seite 114

Advertising
background image

3.2. Remote control commands

798 MPT Titrino

110

Selection of the dialog language.

3.2.2.74. Config.Aux.Set $G

Config.Aux.Set.Date YYYY-MM-DD

Config.Aux.Set.Time HH:MM

Date and time.

Input format of the date: Year-month-day, two-digit, enter leading zeros.

Input format for the time: Hours:minutes, two-digit, enter leading zeros.

Date and time have to be set with &Config.Aux.Set $G just after entry of the

value.

3.2.2.75. Config.Aux.RunNo

0...9999

Current sample number.

Set to 0 on power on and initialization. After 9999, counting starts again at 0.

3.2.2.76. Config.Aux.AutoStart 1...9999, OFF

Number of automatic, internal starts.

3.2.2.77. Config.Aux.StartDelay

0...999 999

Start delay time in s. During this time, the data of the preceding determination

are retained.

3.2.2.78. Config.Aux.ResDisplay bold, standard

Character set for the result display at the end of the determination.

3.2.2.79. Config.Aux.DevName

up to 8 ASCII characters

Name of the instrument for connections with several units. It is advisable to

use only the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and

the numbers 0...9 (ASCII No. 48...57) when the function Setup.AutoInfo

(3.2.2.125) is used at the same time.

If a name has been entered, it will be printed out in the result report (full,

short).

3.2.2.80. Config.Aux.Prog read

only

Output of the program version.

The Titrino sends "798.0010" on requests with $Q.

3.2.2.81. Config.RSSet1 $G

Config.RSSet1.Baud

300, 600, 1200, 2400, 4800, 9600,

19200, 38400, 57600, 115200

Config.RSSet1.DataBit 7,

8

Config.RSSet1.StopBit

1, 2

Config.RSSet1.Parity even,

odd, none

Config.RSSet1.Handsh HWs, SWchar, SWline, none