3.3 Properties of the RS 232 Interface

798 MPT Titrino

128

Titrino as Sender:

Titrino transmission can be stopped by external instruments with XOFF. After

XOFF is received the Titrino completes sending the line already started. If data

output is disabled for more than 6 s by XOFF, E43 appears in the display.

Hardware-Handshake, HWs

Titrino as Receiver :

Titrino as Sender:

The data flow can be interrupted by deactivating the CTS line.

Titrino

external device

Titrino

external device

Titrino

external device

XOFF

XON

Data output

disabled

Data output

enabled

Time

Data output

Data input

RxD

TxD

2

nd

line

L

F

1st line

L

F

DTR

RxD

DTR

RxD

L

F

Time

RTS

CTS

TxD

RTS

CTS

TxD

Time

L

F