16

Adjustment for König

6. Adjustment for König

6.1. Standards

ASTM D 4366

DIN EN ISO 1522

Pendulum and light barrier are adjusted at
delivery. A periodical recertifi cation by authorized
service personal is recommended for equipment
monitoring.

6.2. Duration of pendulum oscillation

According to the Standard one pendulum
oscillation is specifi ed with 1.4 s ± 0.02 s. This
value can be verifi ed by the following procedure:

• Mount the wire release at 6 °

• Raise the lifting table to the stop and check the

zero position.

• Move the pendulum to 6 ° position and fi x it

with the wire release. Press start button.

• Release

pendulum

• Repeat the procedure with the 3° osc. mode. .

After 100 oscillations the elapsed time must be
140 + 2s. If not, please call a service technician
for adjusting.

6.3. Damping time

The damping time is the change of oscillations
from 6° to 3°.The damping time from 6° to 3° must
be 250 ± 10s on the glass plate (for tolerance
please check actual version of the standards.)´This
can be verifi ed with the following procedure:

• Level the instrument with the positioning feet

exactly horizontally control of the bubble level.

• Move the pendulum to 6° position and hold it

with the wire release.