Use open-ended wrench (w/f 17) to remove union nut at hose
connection next to the gas ballast.

Use hex key size 5 to unscrew the handle at one side of the
housing cover.

Use open ended wrench (w/f 15) to turn elbow fitting 1/4 of a
turn, disconnect hose. Do not remove the elbow fitting from
the pump head.
☞ Through reassembly a leak may result.

Position pumping unit at the side, the housing cover at the
side of the inlet block looking upwards.
Use hex key to remove six socket head screws from pump
head and remove upper housing (housing cover, housing cover
insert) and condenser.
☞ Never remove parts by using a spiky or sharp-edged tool

(e. g. screw driver), we recommend to use a rubber mallet
or compressed air (to be blown carefully into port).

Open hose clip at the outlet by using a flat-bladed screw driver
and remove hose nozzle.

Position screw driver as shown in the figure and turn.