Assemble handle.

Use open ended wrench to reconnect hose to elbow fitting.

Tighten union nuts first by hand and then tighten one full turn
using open ended wrench.

Position housing cover.
☞ Move housing cover slightly to make sure that the head

covers are correctly positioned.

Screw in six socket head screws fixing housing cover cross-
wise first slightly, then tighten.
☞ Do not tighten until head cover is in contact with housing,

max. torque 6 Nm.

Position hose clip at the inlet hose of the exhaust waste va-
pour condenser.

Connect the hose to the outlet of the pump.

Close hose clip using a flat pliers.