Notes concerning the operation of the exhaust waste vapour condenser

The

exhaust waste vapour condenser enables an efficient condensation of the pumped vapours at the

outlet.
-

No backflow of condensates.

-

Controlled recovery of condensates.

-

Next to 100% solvent recovery.

-

The isolation cover protects against glass splinters in case of breaking, acts as thermal isolation to
avoid condensation of humidity and is intended to absorb shocks.

Assemble hose nozzles for coolant inlet and coolant outlet pipelines at the exhaust waste vapour
condenser.

Attach the pipelines of the coolant circuit to the respective hose nozzles (hose nozzles 6 mm, see
image) at the waste vapour condenser.

Check hose connections prior to starting operation of the cooling system.

Secure coolant hoses at the hose nozzles (e.g. with hose clip) to prevent their accidental slipping.

The gas outlet (hose nozzle 10 mm) must not be blocked. The exhaust pipeline has always to be free
and pressureless to enable an unhindered discharge of gases.

If necessary connect the exhaust to a suitable treatment plant to prevent the discharge of dangerous
gases and vapours to the surrounding atmosphere.

Attention: Install hoses of the cooling system in a way to avoid flow / dropping of condensed water onto
the pumping unit (especially cables and electronic parts).

Ensure that the coolant outlet pipeline is always free and that it cannot get blocked.

Maximum permissible coolant pressure at the exhaust waste vapour condenser: 6 bar (absolute)
Backflow always pressureless. Assemble a coolant valve always in front of the exhaust waste vapour
condenser (in the inflow).

Comply with the maximum permissible coolant pressures of additional components in the coolant
circuit (e.g cooling water valve).

Avoid overpressure in the coolant circuit (e.g. caused by blocked or squeezed coolant hoses).

Permissible range of coolant temperature at the exhaust waste vapour condenser: -15°C to +20°C

Check the overpressure safety relief device at the exhaust waste vapour condenser regularly, replace
if necessary. Check especially for conglutination and cracks.

Avoid clogging of the exhaust waste vapour condenser due to deposits or freezing solvents, clean if
necessary or select a higher coolant temperature.

In case of condensation: Check liquid level in both catchpots during operation. Do not overfill the
catchpots. Maximum liquid level approx. 80%, to avoid problems when removing the catchpot. Avoid
overflowing of the catchpot.

Check liquid level in both catchpots regularly and drain condensate in time.

Removing the catchpot at outlet:
Remove joint clip, remove catchpot and drain condensate.
Reassemble drained catchpot.

Important: Comply with regulations when disposing solvents/condensates. Reuse if possible, purify if
contaminated.