Using these programs with the sip 10 sipper, Accessories – Hach-Lange LZV935 - DR 6000 Application Software Drinking Water Benutzerhandbuch
Seite 13

English 13
Accessories
•
Spectrophotometer, 436 nm
•
50 mm rectangular cuvette OS
•
Optional SIP 10 sipper: 50 mm pour-through cuvette OS
Interferences
Undissolved substances (turbidity) can lead to cause a high bias result
for color. The interference can be resolved through membrane filtration
(pore width 0.45 µm).
Literature
EN ISO 7887
Procedure for coloration
1.
Select STORED PROGRAMS > SELECT BY NUMBER.
2.
Enter the number 2106 for the program "Coloration
(m
-1
)". Confirm with OK.
Enter the number 2107 for the program "Coloration (PtCo)". Confirm
with OK.
3.
Press START to start the program.
4.
Insert the zero cuvette (distilled water) and press ZERO.
5.
Insert the analysis cuvette with the first sample and press READ.
The result will be displayed on the screen.
Note: In order to analyze further samples, repeat the procedure
from step 5.
Using these programs with the SIP 10 sipper
Information on the installation, configuration, and sample insertion when
using the SIP 10 sipper can be found in the SIP 10 user manual.
Accessories
Description
Order no.
Rectangular cuvettes set (standard deviation = 10 mm,
aligned pair)
2095100
Rectangular cuvette QS, standard deviation = 10 mm
(3.5 mL)
2624410
Rectangular cuvette QS, standard deviation = 50 mm
(17.5 mL)
2624450
Pour-through cuvette QS,
Standard deviation = 3 mm, center height = 10 mm,
height = 40 mm
LZV638
Pour-through cuvette QS,
Standard deviation = 50 mm, center height = 10 mm,
height = 40 mm
LZV649
Pour-through cuvette QS,
Standard deviation = 10 mm, center height = 10 mm,
height = 40 mm
LZV510
SIP 10 sipper module set for DR 6000 complete with tray,
tube set and 1 cm quartz glass pour-through cuvette, EU
LQV157.99.30001
Tube set for DR 6000 "drinking water application,"
complete, including Pharmed tubes, pump tube (Lagoprene
pre-installed with cover and connections), stainless steel
tube adapter
LZQ100