WIKA operating instructions model DIH10-Ex
13
3330872.01 02/2014 GB/D
Explosion protection, power supply
Only for connection to a certified intrinsically safe circuit with the following
maximum values:
Maximum terminal voltage U
0
DC 30 V
Maximum current l
0
100 mA
Maximum power P
0
1 W
Maximum capacitance C
i
negligibly small
Maximum inductance L
i
negligibly small
Ambient conditions
Operating temperature
Standard version: 0 ... 80 °C
Ex version: 0 ... 60 °C
Storage temperature
-20 ... +80 °C
Module housing
Dimensions
Ø 43.0 / 48.5 / 62.5 mm x 37 mm
Case material
Polycarbonate
Ingress protection
Front: IP 67
Connection: IP 20 or IP 00
Weight
approx. 55 g
Connection
Plug-in terminal strip, up to 1.5 mm²
open cable ends
Programmable features
Scale range, display time, decimal point,
unit (°C/°F), zero point stabilisation,
programming lock, reference points, TAG
number
Case designs
(see chapter 4.2)
■
Connection head (standard)
■
Wall-mount case
■
Case with pipe support
CE conformity
EMC directive
2004/108/EC, EN 61326 emission
(group 1, class A) and interference
immunity (industrial application)
ATEX directive
94/9/EC, II 2 G Ex ia IIC T6
For further specifications see WIKA data sheet AC 80.11 and the order
documentation.
3. Specifications