Transport, scope of delivery, installation, Notice, 5 installation of the machine – Retsch AS 450 basic Benutzerhandbuch

Seite 14

Advertising
background image

Transport, scope of delivery, installation

14

NOTICE

Ambient temperature

Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.

Do not exceed or fall below the permitted temperature range of the
machine (5°C to 40°C / ambient temperature).

Pos : 7.13 /00003 Standard Kapitel/General M odul Luftfeuchtigkeit @ 0\mod_1228918538693_9.doc @ 4738 @ @ 1

Atmospheric humidity:

Maximum relative humidity 80% at temperatures up to 31°C, decreasing linearly up
to 50% relative humidity at 40°C

Pos : 7.14 /00004 Warnhinweis e/H 0011 HIN WEIS Luftfeuc htig keit @ 0\mod_1228918882628_9.doc @ 4788 @ @ 1

NOTICE

Atmospheric humidity

Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.

Do not exceed the admissible range for atmospheric humidity.

Pos : 7.15 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Aufs tell en des Ger ätes @ 0\mod_1226498849756_9.doc @ 3464 @ 2 @ 1

4.5 Installation of the machine

Pos : 7.16 /00010 Bedi enungsanleitung en Kapitels ammlungen/AS450_basic /0006 Transport, Li eferumfang und Aufstellen /AS450 basic M odul Ger ät ausrichten @ 4\mod_1329993637090_9.doc @ 26800 @ @ 1

The device should be erected on a level and stable surface which will not sag and
which is suitable for use during operation.

A level surface ensures the even distribution of the sample across the sieves and
stability of the machine.

Fig. 1: Aligning the device

The flexible feet underneath the machine enables it to be adjusted to the surface.

Unscrew the lock nut (KM).

Twist the flexible foot of the device (GF) out or in as necessary.

Tighten the lock nuts on the feet.

Pos : 7.17 /00003 Standard Kapitel/General M odul Aufstellungshöhe @ 0\mod_1228918538349_9.doc @ 4724 @ @ 1

Installation height: maximum 2000 m above sea level

Pos : 7.18 /00004 Warnhinweis e/H 0058 HIN WEIS Boden Aufstellen Vibr ationen Boden (AS450 basi c) @ 4\mod_1332943698568_9.doc @ 27538 @ @ 1

NOTE

3.H0004

Erection

Vibrations can occur depending on the operating status of the device.

Only erect the device on a level, flat and horizontal surface. The base
must be stable and must not vibrate.

Pos : 7.19 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Spannstange eins etzen @ 4\mod_1330075289191_9.doc @ 26818 @ 2 @ 1

KM

GF

Advertising