Transport, scope of delivery, installation

14

NOTICE

Ambient temperature

–

Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.

•

Do not exceed or fall below the permitted temperature range of the
machine (5°C to 40°C / ambient temperature).

Pos : 7.13 /00003 Standard Kapitel/General M odul Luftfeuchtigkeit @ 0\mod_1228918538693_9.doc @ 4738 @ @ 1

Atmospheric humidity:

Maximum relative humidity 80% at temperatures up to 31°C, decreasing linearly up
to 50% relative humidity at 40°C

Pos : 7.14 /00004 Warnhinweis e/H 0011 HIN WEIS Luftfeuc htig keit @ 0\mod_1228918882628_9.doc @ 4788 @ @ 1

NOTICE

Atmospheric humidity

–

Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.

•

Do not exceed the admissible range for atmospheric humidity.

Pos : 7.15 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Aufs tell en des Ger ätes @ 0\mod_1226498849756_9.doc @ 3464 @ 2 @ 1

4.5 Installation of the machine

Pos : 7.16 /00010 Bedi enungsanleitung en Kapitels ammlungen/AS450_basic /0006 Transport, Li eferumfang und Aufstellen /AS450 basic M odul Ger ät ausrichten @ 4\mod_1329993637090_9.doc @ 26800 @ @ 1

The device should be erected on a level and stable surface which will not sag and
which is suitable for use during operation.

A level surface ensures the even distribution of the sample across the sieves and
stability of the machine.

Fig. 1: Aligning the device

The flexible feet underneath the machine enables it to be adjusted to the surface.

•

Unscrew the lock nut (KM).

•

Twist the flexible foot of the device (GF) out or in as necessary.

•

Tighten the lock nuts on the feet.

Pos : 7.17 /00003 Standard Kapitel/General M odul Aufstellungshöhe @ 0\mod_1228918538349_9.doc @ 4724 @ @ 1

Installation height: maximum 2000 m above sea level

Pos : 7.18 /00004 Warnhinweis e/H 0058 HIN WEIS Boden Aufstellen Vibr ationen Boden (AS450 basi c) @ 4\mod_1332943698568_9.doc @ 27538 @ @ 1

NOTE

3.H0004

Erection
–

Vibrations can occur depending on the operating status of the device.

•

Only erect the device on a level, flat and horizontal surface. The base
must be stable and must not vibrate.

Pos : 7.19 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Spannstange eins etzen @ 4\mod_1330075289191_9.doc @ 26818 @ 2 @ 1

KM

GF