Cleaning and service
32
The AS450 basic is protected by a device fuse .
The external fusing must be T5.0A (230V) F15A (110V).
•
Turn the device off at the mains (H).
•
Pull the mains plug from the low heat device socket (J).
•
Unscrew the fuse insert (I) and remove the fuse (IS).
CAUTION
A fuse must be replaced by one of the same type and with the same fuse value.
•
Place the new fuse in the fuse insert.
NOTE
Do not tighten the fuse insert too much when screwing the fuse in!
•
Screw the fuse insert (I) back in.
NOTE
If the fuse triggers again immediately after being replaced, there is a fault in the
device which must be rectified.
There are two fuses: one in the live conductor and one in the neutral conductor of
the power supply.