•Calibration

General
The URT/R is factory calibrated and you can select any input range
without recalibrating the unit. However, if you change the RTD type,
you must calibrate the inputs.
Allow the instrument 1 minutes of powered operation (to reach a
stable temperature) before calibration and, during calibration, allow a
few seconds for signals to stabilise before accepting the value.
Equipment requirements
• A suitable RTD simulator (accurate to 0.0 °C)

Inputcalibrationpoints

Calibrationpoint

Pt100

Ni120

Display

Temperature(°C)

Display

Temperature(°C)

1

-200

-200.0 °C

-80n

-80.0 °C

2

700P

700.0 °C

320n

320.0 °C

3

-100

-100 °C

-80n

-80.0 °C

4

200P

200.0 °C

200n

200.0 °C

Terminalconnectionsforinputcalibration

CalibrationStage

Signaltype

Terminal

RTD Inputs

RTD Simulator A

2

RTD Simulator B

3

RTD Simulator Bsense

4