Calibration Procedure with Taber Abrasion Holder
• Calibrate the instrument without the Taber Abrasion Holder.
• Install the Taber Abrasion Holder in front of the sphere entrance.
• Calibrate the instrument once more, but only dark- and bright calibration.
When you are asked to calibrate Clarity, please press the ”set-up“ button.
• Perform measurements without a sample in Haze, Transmission and Clarity.
The result should be: H = 0.00, T = 100 and C = 100.
If the Taber Abrasion Holder is removed, calibrate the instrument again.
Note:
When working with the Taber Abrasion Holder, only measure Haze or Transmission.
Clarity measurements are not possible with the holder due to changed