Control elements and operation – ALTANA Pendulum Hardness Tester Manual Benutzerhandbuch
Seite 15

14
Control elements and operation
5. Control elements and operation
5.1. Control elements
Four buttons to control the instrument sit on the
front panel
start/stop
signal
modus
enter
To prepare a measurement fi rst choose the
appropriate method by pressing the ”mode“
button: four settings are available:
• Persoz 4° time
• Koenig 3° osc.
• No Std. 4° osc.
• No Std. 3° time
According to the standards for
• Persoz: oscillation time in seconds
• König: number of oscillations
The combinations
• 4° amplitude / number of oscillations and
3° amplitude / oscillation time are not defi ned
in the standards.
• The selected method is confi rmed by pressing
the ”enter”-button.
• Move the pendulum with the displacement arm
to the starting position.
• Press the wire release to hold the pendulum in
position
• Press “Start/Stop“- button, an asterisk appears
on the display and the red LED lights up
(no air draft or vibration from now on)
English