4.1 Zero Adjustment


This function will auto zero the sensor signal. Make sure to perform zero adjustment only in a
clean environment without combustible gases present.

??Switch the function switch to “TEST”- position. Instrument will perform an automatic

selftest.

??Press “Cal” using the keyboard pin. The display will show “000” indicating the automatic

zero adjust.

??When the zero adjust is finished, display will show “GCAL”.
??Now you can call up additional functions or leave the service mode by turning the

function switch to the “ON” or “OFF”- position.


4.2 Span Adjustment


The sensitivity of the sensor may decrease with time due to exposure to substances poisoning
the catalytic bead. Therefore it is important to recalibrate the Compur Ex plus in regular
intervales.

Items needed for calibration:

-

Cal gas adapter

(see accessories)

-

Valve with flow meter

-

Tubing (Silicon free)

-

Span gas

(we recommend to use
span gas with 50 % LEL)




















Ex plus

Methan

%UEG/LEL

ON

TEST

OFF

Fct

Cal