Conduct this performance test daily
Do not conduct performance test by picking up baby out of bed and waiting for the alarm to sound
• The performance test should be conducted while baby is not in bed.
• Turn device on and take two steps back from bed.
• Ensure that alarm goes off within 20 seconds (due to absence of breathing movements).
• The alarm may sound with or without preliminary ticking.
• Check that alarm can be heard in all locations throughout the home or place of care.
• In case that alarm does not sound within 20 seconds refer to troubleshooting at page .
• Performance test must be conducted daily.

False alarms
Isolated cases of false alarms are possible. Check if caused by faulty connection of the sensor
pad(s) to the control unit, connection of only one sensor pad when baby can move off the immedia-
te area, connection of two sensor pads far apart from each other when baby not crawling yet, or
failure of the sensor pads to be directly against the mattress.

In case of false alarms, kindly refer to troubleshooting guide. If the problem cannot be solved, plea-
se contact your nearest BABYSENSE representative.

In the event of alarm
Act immediately according to the following:
1. Approach the bed and check that baby is breathing. Note pallor (color of skin), muscle tone

and movement.

2. If baby does not seem normal, pinch baby's heel and check that there is no cause for choking,

such as an object lodged in baby’s throat.

When breathing resumes, consult a doctor.

3. If breathing does not resume, apply first aid and seek medical assistance immediately.

Caring for your babysense
"Babysense" requires no special treatment beyond replacing batteries when the battery indicator
light goes on. The sensor pads and control unit may be wiped with a damp cloth.
Batteries should be removed and the device stored in a dry shaded place when not in use.

a

It is recommended to replace sensors every two years.

5 Appendix

Troubleshooting Guide
Should problems arise with the telephone, please refer to the following information first.
In the case of technical problems, contact our Hotline. Please refer to our website
www.audioline.de for the telephone number.
In the case of claims under the terms of guarantee, contact your sales outlet. There is a 2 year