Technical data – Beurer ME80 Benutzerhandbuch
Seite 47

47
Problem
Possible cause
solution
On launching the program,
the message “Verification
error” is displayed.
The device is not con-
nected to a PC or the
Beurer ECG Manager did
not install correctly.
Make sure that the ME 80 is
connected to your computer
and restart the program. If
the verification still fails, read
the information on installation
on page 12 and reinstall the
program.
9. technical Data
Model
ME 80
System of measuremen
single-channel ECG in freely selectable frontal posi-
tions/ECG signal referenced to ground (earth)
Bandwidth
0.05 to 40 Hz
Sample rate
256 Hz
Measurement range for heart
rate:
5 to 199 beats per min.
Memory
99 measurements
Power supply
Integrated, rechargeable lithium battery
Battery capacity
Over 1,000 measurements on a full charge
Transport conditions
-20°C to +60°C, 10%–95% humidity
Storage conditions
-20°C to +60°C, 10%–95% humidity
Operating conditions
+5°C to +40°C, 30%–85% humidity
Weight
40 g