l

[->] key to select:

- menu change

- list entry

-

Select editing position

l

[+] key:

-

Change value of a parameter

l

[ESC] key:

- interrupt input

-

jump to the next higher menu

Adjustment system

The sensor is adjusted via the four keys of the

indicating and adjustment module. The LC display

indicates the individual menu items. The functions

of the individual keys are shown in the above

illustration. Approx. 10 minutes after the last

pressing of a key, an automatic reset to measured

value indication is triggered. Any values not

confirmed with [OK] will not be saved.

5.4 Setup procedure

Parameter adjustment

To carry out this adjustment, the distance is

entered for full and empty vessel. If these values

are not known, an adjustment with distances, for

example, of 10 % and 90 % is also possible.

Starting point for these distance specifications is

the lower edge of the flange (flange versions), for

all other versions the lower edge of the transducer.
The real product level during this adjustment is not

important, because the min./max. adjustment is

always carried out without changing the product

level. These settings can be made ahead of time

without the instrument having to be installed.
In the main menu item "Basic adjustment", the

individual submenu items should be selected one

after the other and provided with the correct

parameter values.

Carrying out min. adjustment

Proceed as follows:
1 Move from the measured value display to the

main menu by pushing [OK].

> Basic adjustment

Display
Diagnostics
Service
Info

2 Select the menu item "Basic adjustment" with

[->] and confirm with [OK]. Now the menu item

"Min. adjustment" is displayed.

Min. adjustment
0.00 %
=
5,000 m(d)

4,000 m(d)

3 Prepare the % value for editing with [OK] and

set the cursor to the requested position with [-

>]. Set the requested percentage value with [+]

and save with [OK]. The cursor jumps now to

the distance value.

4 Enter the appropriate distance value in m

(corresponding to the percentage value) for the

empty vessel (e.g. distance from the sensor to

the vessel bottom).

5 Save the settings with [OK] and move to "Max.

adjustment" with [->].

Carrying out max. adjustment

Proceed as follows:

Max. adjustment
100.00 %
=
1,000 m(d)

2,000 m(d)

1 Prepare the % value for editing with [OK] and

set the cursor to the requested position with [-

>]. Set the requested percentage value with [+]

and save with [OK]. The cursor jumps now to

the distance value.

2 Enter the appropriate distance value in m

(corresponding to the percentage value) for the

full vessel. Keep in mind that the max. level

must lie below the dead band.

3 Save the settings with [OK] and move to

"Medium selection" with [->].

Medium selection

Each product has different reflective properties. In

addition, there are various interfering factors which

have to be taken into account: agitated product

surfaces and foam generation (with liquids); dust

VEGASON S 61

29

Set up with the indicating and adjustment module