Trouble shooting – BrandTech Scientific Macro Transferpettor Benutzerhandbuch

Seite 22

Advertising
background image

– 22 –

Problem

Possible Cause

Corrective Action

Pipetted volume too

low/Pipetting key cannot

be pushed fully down to

the end plate.

The tip has been installed too

far into the tip mount. Piston

rod is loose.

- Remove the tip (= Transferpettor-cap).

"Changing the Tip", page 12.

- Tighten the piston rod locking device.

"To fit the new piston rod", page 17.

- Attaching the tip, page 13.

Pipette leaks.

The tip does not fit properly.

Damaged seal.

- Reinstall the tip (= Transferpettor-cap) and slide it

on again.

"Changing the Tip", page 12.

- Change the seal, page 14.

Air bubbles in the liquid

drawn in.

The liquid was drawn in too

quickly.

The tip does not fit properly.

Damaged seal.

- Let the pipetting key slide back slowly.

- Reinstall the tip (= Transferpettor-cap) and slide it

on again.

"Changing the Tip", page 12.

- Change the seal, page 14.

Liquid above the seal.

Damaged tip or seal.

- Change tip (= Transferpettor-cap) or seal.

"Changing the Tip", page 12.

"Changing the Seal", page 14.

Liquid remaining in the

tip.

Tip not properly seated.

Piston rod is loose.

Damaged seal.

- Remove the tip (= Transferpettor-cap).

"Changing the Tip", page 12.

- Tighten the piston rod locking device.

"To fit the new piston rod", page 17.

- Fit the tip, page 13.

- Change the seal, page 14.

Trouble Shooting