WIKA operating instructions model CPG1000

11568047.02 08/2012 GB/D

GB

2. Safety

DANGER!

Danger of death caused by electric current

Operation using a defective power supply unit (e.g. short circuit from the

mains voltage to the output voltage) can result in life-threatening voltages

at the instrument!

Special conditions for safe use

WARNING!

If the CPG1000 has been exposed to overload or sudden physical shock

(e.g. being dropped) it should be checked for signs of damage that may

cause concerns about safety. If there is any doubt, return the unit to WIKA

for evaluation. For information on returning the instrument see chapter

9.2 "Return".

The CPG1000 is not intended for use with flammable substances and must

be used only in locations which provide sufficient protection against the

intrusion of solid foreign objects or water which could impair the safety.

CAUTION!

To avoid possible damage to the calibration reference or to equipment

under test:

If the message "OL" is displayed, the measuring range has been

exceeded and the pressure source must be removed immediately from

the CPG1000 in order to prevent damage to the internal sensor.

The maximum permitted torque of the thread is 13.5 Nm = 10 ftlbs. This

permitted torque must NEVER be exceeded.

When screwing the NPT thread of the CPG1000 to an adapter or

a process connection, it is necessary to use an additional sealing

compound, e.g. Teflon tape between the threads.

The sealing has to be made with Teflon tape directly via the NPT thread

and not via the manifold at the CPG1000.