INFICON SKY CDGxxxD (RS232C) Benutzerhandbuch

Seite 9

Advertising
background image

tira49d1 (2009-01) RS232C.kal

9

(abgeschlossen)

Parameter Name

Daten Typ

Beschreibung

Byte Nr. 2

Byte Nr. 3

Bemerkung

Extended Error

uint8_T / Ro

L-Byte

54

Bit 0

Atm.-Pressure out of range

L-Byte

Bit 1

Temperature out of range

Bit

2

Reserve

Bit

3

Reserve

Bit 4

Cal-Mode falsch

Bit 5

Pressure Underflow

Bit 6

Pressure Overflow

Bit

7

Zero-Adjust-Overflow

Extended Error

uint8_T / Ro

H-Byte

55

Bit 0

PT1000 fault

4)

H-Byte

Bit 1

Heaterblock overtemp.

4)

Bit

2

Reserve

Bit

3

Reserve

Bit

4

Reserve

Bit

5

Reserve

Bit

6

Reserve

Bit

7

Reserve

Pressure range

uint8_T / Ro

56

0

F.S.R. = E-3

(Exponent)

1

F.S.R.

=

E-2

2

F.S.R.

=

E-1

3

F.S.R. = E 0

4

F.S.R.

=

E+1

5

F.S.R.

=

E+2

6

F.S.R.

=

E+3

7

F.S.R.

=

E+4

Pressure range

uint8_T / Ro

57

0

Mantisse = 1.0

(Mantisse)

1

Mantisse = 1.1

2

Mantisse = 2.0

3

Mantisse = 2.5

4

Mantisse = 5.0

Gauge config

uint8_T / Ro

L-Byte

58

0

= Analogout 0 … 10.24 V

1

= Analogout 1 … 9 V

CDG type

uint8_T / Ro

L-Byte

59

0

= CDG025D

1

=

CDG045D

2

=

CDG100D

3

=

CDG160D

4

=

CDG200D

RW = Read / Write

Ro = Read only

1)

Werkseinstellung

2)

Umrechnung

→ Kapitel 1.1, Byte 6 "Lesebefehl"

3)

Umrechnung

→ Kapitel 1.1, Byte 4 und 5 "Druckeinheit"

4)

Nur CDG045D, CDG100D, CDG160D, CDG200D