30
5 Configuration and parameter tables
Additional parameters
Para-
meter
Explanation
Value range factory
setting
X
C236 Frequency at
range end
10 —
1000Hz
1000
Para-
meter
Explanation
Value range factory
setting
X
SCL
Scaling start value
-1999 to
+9999 digit
0
SCh
Scaling end value
-1999 to
+9999 digit
100
DF
Filter time constant
0.0 —
100.0sec
0.6
R FE
Total resistance for
resistance transmitters
0 — 4000
Ω
1000
C001 Simulation of
measurement output
0 — 110%
(111 =
switched off)
111