2 description of the remote control commands – Metrohm 758 KFD Titrino Benutzerhandbuch

Seite 103

Advertising
background image

3.2 Remote control commands

758 KFD Titrino

99

3.2.2 Description of the remote control commands

3.2.2.1. Mode

$G, $S, $H, $C

Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and

continue with $C.

$G also serves to continue after inquiries of identifications and sample size af-

ter the start (see 3.2.2.28) as well as after inquiries of calibration temperature

and pH values of buffers (see 3.2.2.39 and 3.2.2.40).

3.2.2.2. Mode.QuickMeas $G,

$S

Start and stop of a measurement in the basic mode with the parameters

(measured quantity, measuring input) of the current method. Corresponds to

the <meas/hold> key. In TIP the measured quantity is selected with

&Mode.Parameter, see 3.2.2.47.

With an ongoing measurement, the current mode can be started. This stops

the measurement automatically.

3.2.2.3. Mode.Select KFT, SET, MEAS, CAL, TIP

Mode.SETQuantity pH, U, Ipol, Upol

Mode.MEASQuantity pH, U, Ipol, Upol, T

Mode.KFTQuantity Ipol, Upol

Selection of the standard mode. Mode and the measured quantity belong to

the complete selection.

If a method is selected from the method memory, the nodes &Mode.Select

and &Mode.XXXQuantity are overwritten with mode and measured quantity of

the corresponding user method.

3.2.2.4. Mode.Name read

only

Name of the current method in the working memory. $Q sends 8 ASCII char-

acters. Standard methods carry the name ********. The node can be set

read + write, see 3.2.2.105.


3.2.2.5. Mode.Def.Formulas.1.Formula

EPX, CXX, RSX, +, –, ∗, /, (, )

Mode.Def.Formulas.1.TextRS

up to 8 ASCII characters

Mode.Def.Formulas.1.Decimal 0...2...5

Mode.Def.Formulas.1.Unit

up to 6 ASCII characters

Mode.Def.Formulas.1.Limits ON,

OFF

Mode.Def.Formulas.1.LoLim

0...±999 999

Mode.Def.Formulas.1.UpLim

0...±999 999

Mode.Def.Formulas.1.Output active,

pulse,

OFF

Mode.Def.Formulas.2.Formula

etc. up to .9

Entry of formulas. Rules for formula entry, see page 42.
Example: "(EP2–EP1)∗C01/C00"

In addition to the formula, a text for result output, the number of decimal

places and a unit for the result output can be selected. "No unit" is selected

with the blank string.

Advertising