Gossen DIGISKY Benutzerhandbuch

Seite 18

Advertising
background image

GOSSEN Foto- und Lichtmesstechnik GmbH

18

6.2.2

Flash Light Reading

– Analysis

Use the ring controller ...

to select synchronization

speed and analyze flash

component.

Flash component is

displayed as a percentage.







































i

Please note that some flash units must first be triggered once when set to lower power levels.
Stored power is still at the higher level, resulting in erroneous measurements.

i

The ratio between flash and ambient light can be influenced by changing synchronization
speed. This is interesting for brightening up the image with flash, or when ambient light is not
desired.

!

Warning in Case of User Error
A warning may be caused by any of the following conditions.

General:

Measurement has been performed outside of the measuring
range. There are no usable results

– EV display: OVR. Flash

was too bright or too weak. Change conditions and measure
again.
See also Technical Data

Measuring Range on page 31.

NonCord

Time out

– Flash was not triggered during “flash standby time”.

Break

– Flash standby time was stopped by pressing the

measurement key.

Cord:

The synchronizing cable was not connected.
The synchronizing cable is defective and must be replaced.

Radio:

No flash regardless of setting:

Check menu settings.
RF module - RF channel on page 26.

Flash is only triggered in the “All” setting.

The menu settings are correct

– please select the right

group.

Advertising