5 measuring 5.3 setup at the controller – Henkel LOCKTITE 97211 Benutzerhandbuch

Seite 20

Advertising
background image

20

5 Measuring

5.3 Setup at the Controller

• Open the main directory PERIPHERY SETUP, see operating manual of the used

controller.

• Set the directory point Flow.Monitor respectively Disp.Control from [ O ] to [ I ]. The

subdirectory is open.

• Choose the required tolerance and

store it.

Toler.: low [ 0 ] / [ 1 ]
Toler.: middle [ 1 ] / [ 0 ]
Toler.: high [ 0 ] / [ 1 ]
Tolerance 35%

The values of these tolerance classes are: Toler.: low 15%
Toler.: middle 35%
Toler.: high 55%

Start with the tolerance middle (factory setting). Dispense several times to see, if it is too
sensitive or too insensitive, e. g. do you get too many error messages due to normal
fluctuations in the product or no error message at all.

If you get too many error messages, set tolerance to high or increase the tolerance to a
value, which is big enough that the relationship between ok and bad dispensing
correspond to your requirements.

If you get too little error messages, set tolerance to low or decrease the tolerance to a
value, which is small enough that the relationship between ok and bad dispensing
correspond to your requirements.



At the Automatic Controller 97103 the tolerance values low, middle and high are
available only.

Advertising